• Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900
  • Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900
  • Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900
  • Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900
  • Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900
  • Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900

Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900

Warranty: One Year
Display: Digital
Portable: Portable
Appliance: Food Analysis, Medical Research, Biochemistry
Certification: ISO9001
Transport Package: Standard Export Package
Samples:
US$ 780/Piece 1 Piece(Min.Order)
| Request Sample
Gold Member Since 2024

Suppliers with verified business licenses

Guangxi, China
Importers and Exporters
The supplier has import and export rights
High Repeat Buyers Choice
More than 50% of buyers repeatedly choose the supplier
Experienced Team
The supplier has 5 foreign trading staff(s) and 4 staff(s) with over 6 years of overseas trading experience
In-stock Capacity
The supplier has In-stock capacity
to see all verified strength labels (12)

Basic Info.

Model NO.
Bante900
Specification
30mx25mx200mm
Origin
China

Product Description

Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900Bante900 Benchtop Multiparameter Water Quality Meter

High-precision multiparameter water quality meter contains measurement modes for pH, ORP, ion concentration, conductivity, TDS, salinity, resistivity and dissolved oxygen, the setup menu offers 17 optional settings for customization.

pH Mode:
• Multiparameter water quality meter is equipped with a large backlit LCD display.
• 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers.
• Automatic electrode diagnosis shows the pH slope and zero offset.

ORP Mode:
• 1 point offset calibration allows adjusting the displayed value to a known standard.
• Relative and absolute millivolt modes ensure the reliable ORP measurements.

Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900Ion Concentration Mode:
• 2 to 5 points calibration, including the selection of 8 concentration points.
• Automatic electrode diagnosis shows the calibration points and slopes.
• Direct ion concentration readout simplifies the measurement process.
• Selectable concentration units (ppm, mg/L, mol/L) and ionic valency.

Conductivity/TDS/Salinity/Resistivity Modes:
• 1 to 5 points calibration with automatic recognition for conductivity standards.
• Selectable cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS factor, linear and
  pure water compensations, seawater and practical salinity measurement modes.

Dissolved Oxygen Mode:
• 1 or 2 points calibration using the air-saturated water or zero oxygen solution.
• Salinity and barometric pressure compensations eliminate the measurement errors.
Bante900 Benchtop Multiparameter Water Quality Meter Bante pH Meter 900
Other Features:
• Automatic temperature compensation ensures accurate readings over the entire range.
• Auto-read function senses and locks the measurement endpoint.
• Calibration due alarm reminds the user to calibrate the meter regularly.
• Setup menu allows setting the number of calibration points, stability criteria, temperature unit,
  date and time, auto-power off, etc.
• Reset function automatically resumes all settings back to the factory defaults.
• Expanded memory stores or recalls up to 500 data sets.
• USB communication interface for data transfer and timed interval readings.


 

 Model

 

Bante900

   pH

   

   Range

 

-2.000 to 20.000 pH

   Resolution

 

0.001 / 0.01 / 0.1 pH, selectable

   Accuracy

 

±0.002 pH

   Calibration

 

1 to 5 points

   pH Buffer Options

 

USA / NIST / DIN / 2 custom buffers

   ORP

   

   Range

 

-1999.9 to 1999.9 mV

   Resolution

 

0.1 / 1 mV, selectable

   Accuracy

 

±0.2 mV

   Calibration

 

1 point

   Measurement Modes

 

Relative / Absolute mV

   Ion Concentration

   

   Range

 

0.00 to 19999 (depending on the range of ISE)

   Resolution

 

0.001, 0.01, 0.1, 1, automatic

   Accuracy

 

±0.5% F.S. (monovalent), ±1% F.S. (divalent)

   Calibration

 

2 to 5 points

   Calibration Solutions

 

0.001 / 0.01 / 0.1 / 1 / 10 / 100 / 1000 / 10000

   Measurement Units

 

ppm, mg/L, mol/L, mmol/L

   Conductivity

   

   Range

 

0 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm

   Resolution

 

0.001, 0.01, 0.1, 1, automatic

   Accuracy

 

±0.5% F.S.

   Calibration

 

1 to 5 points

   Calibration Solutions

 

10 µS / 84 µS / 1413 µS / 12.88 mS / 111.8 mS

   Temperature Coefficient

 

0.0 to 10.0%/°C

   Pure Water Compensation

 

Yes

   Reference Temperature

 

20 / 25°C

   Cell Constant  

 

K=0.1 / 1 / 10 / Custom

   TDS

   

   Range

 

0 to 10.00, 100.0, 1000 ppm, 10.00, 200.0 ppt

   Resolution

 

0.01, 0.1, 1, automatic

   Accuracy

 

±1% F.S.

   TDS Conversion Factor

 

0.1 to 1.0 (default 0.5)

   Salinity

   

   Range

 

0.00 to 80.00 ppt / 0.00 to 42.00 psu

   Resolution

 

0.01

   Accuracy

 

±1% F.S.

   Resistivity

   

   Range

 

0.00 to 20.00 MΩ

   Resolution

 

0.01

   Accuracy

 

±1% F.S.

   Dissolved Oxygen

   

   Range

 

0.00 to 20.00 mg/L / 0.0 to 200.0 % saturation

   Resolution

 

0.01 / 0.1, selectable

   Accuracy

 

±0.2 mg/L / ±2.0 %

   Calibration

 

1 or 2 points

   Pressure Correction

 

60.0 to 112.5 kPa / 450 to 850 mmHg, manual

   Salinity Correction

 

0.0 to 50.0 g/L, manual

   Temperature

   

   Range

 

0 to 105°C / 32 to 221°F

   Resolution

 

0.1°C/°F

   Accuracy

 

±0.5°C / ±0.9°F

   Calibration

 

1 point

   Calibration Range

 

Reading ±10°C/°F

   Other Specifications

   

   Temperature Compensation

 

0 to 100°C / 32 to 212°F, manual or automatic

   Hold Function

 

Manual / Auto-read measurement endpoint

   Stability Criteria

 

Low / High

   Calibration Due Alarm

 

1 to 31 days or off

   Memory

 

Stores up to 500 data sets

   Output

 

USB communication interface

   Connector

 

BNC, 6-pin mini-DIN, 3.5 mm jack socket

   Display

 

Custom LCD (125 × 100 mm)

   Power Supply

 

5V DC power adapter

   Environmental Conditions

 

0 to 50°C / 32 to 122°F, relative humidity < 80%

   Dimensions

 

210 × 188 × 60 mm

   Weight

 

1.5 kg


 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now