High-precision multiparameter water quality meter contains measurement modes for pH, ORP, ion concentration, conductivity, TDS, salinity, resistivity and dissolved oxygen, the setup menu offers 17 optional settings for customization.
pH Mode:
• Multiparameter water quality meter is equipped with a large backlit LCD display.
• 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers.
• Automatic electrode diagnosis shows the pH slope and zero offset.
ORP Mode:
• 1 point offset calibration allows adjusting the displayed value to a known standard.
• Relative and absolute millivolt modes ensure the reliable ORP measurements.
Model |
Bante900 |
|
pH |
||
Range |
-2.000 to 20.000 pH |
|
Resolution |
0.001 / 0.01 / 0.1 pH, selectable |
|
Accuracy |
±0.002 pH |
|
Calibration |
1 to 5 points |
|
pH Buffer Options |
USA / NIST / DIN / 2 custom buffers |
|
ORP |
||
Range |
-1999.9 to 1999.9 mV |
|
Resolution |
0.1 / 1 mV, selectable |
|
Accuracy |
±0.2 mV |
|
Calibration |
1 point |
|
Measurement Modes |
Relative / Absolute mV |
|
Ion Concentration |
||
Range |
0.00 to 19999 (depending on the range of ISE) |
|
Resolution |
0.001, 0.01, 0.1, 1, automatic |
|
Accuracy |
±0.5% F.S. (monovalent), ±1% F.S. (divalent) |
|
Calibration |
2 to 5 points |
|
Calibration Solutions |
0.001 / 0.01 / 0.1 / 1 / 10 / 100 / 1000 / 10000 |
|
Measurement Units |
ppm, mg/L, mol/L, mmol/L |
|
Conductivity |
||
Range |
0 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm |
|
Resolution |
0.001, 0.01, 0.1, 1, automatic |
|
Accuracy |
±0.5% F.S. |
|
Calibration |
1 to 5 points |
|
Calibration Solutions |
10 µS / 84 µS / 1413 µS / 12.88 mS / 111.8 mS |
|
Temperature Coefficient |
0.0 to 10.0%/°C |
|
Pure Water Compensation |
Yes |
|
Reference Temperature |
20 / 25°C |
|
Cell Constant |
K=0.1 / 1 / 10 / Custom |
|
TDS |
||
Range |
0 to 10.00, 100.0, 1000 ppm, 10.00, 200.0 ppt |
|
Resolution |
0.01, 0.1, 1, automatic |
|
Accuracy |
±1% F.S. |
|
TDS Conversion Factor |
0.1 to 1.0 (default 0.5) |
|
Salinity |
||
Range |
0.00 to 80.00 ppt / 0.00 to 42.00 psu |
|
Resolution |
0.01 |
|
Accuracy |
±1% F.S. |
|
Resistivity |
||
Range |
0.00 to 20.00 MΩ |
|
Resolution |
0.01 |
|
Accuracy |
±1% F.S. |
|
Dissolved Oxygen |
||
Range |
0.00 to 20.00 mg/L / 0.0 to 200.0 % saturation |
|
Resolution |
0.01 / 0.1, selectable |
|
Accuracy |
±0.2 mg/L / ±2.0 % |
|
Calibration |
1 or 2 points |
|
Pressure Correction |
60.0 to 112.5 kPa / 450 to 850 mmHg, manual |
|
Salinity Correction |
0.0 to 50.0 g/L, manual |
|
Temperature |
||
Range |
0 to 105°C / 32 to 221°F |
|
Resolution |
0.1°C/°F |
|
Accuracy |
±0.5°C / ±0.9°F |
|
Calibration |
1 point |
|
Calibration Range |
Reading ±10°C/°F |
|
Other Specifications |
||
Temperature Compensation |
|
0 to 100°C / 32 to 212°F, manual or automatic |
Hold Function |
|
Manual / Auto-read measurement endpoint |
Stability Criteria |
|
Low / High |
Calibration Due Alarm |
|
1 to 31 days or off |
Memory |
|
Stores up to 500 data sets |
Output |
|
USB communication interface |
Connector |
|
BNC, 6-pin mini-DIN, 3.5 mm jack socket |
Display |
|
Custom LCD (125 × 100 mm) |
Power Supply |
|
5V DC power adapter |
Environmental Conditions |
|
0 to 50°C / 32 to 122°F, relative humidity < 80% |
Dimensions |
|
210 × 188 × 60 mm |
Weight |
|
1.5 kg |