TECHNICAL CONFIGURATION LIST, STANDARD | ||||
NO | Description | Qty | Description | Note |
I. | Machine | 1 | NEW STANDARD 322Machine | |
1 | NEW STANDARD 322Base machine and Control System | |||
1 | Manual Zoom 0.7X-4.5X Manual zoom (include Coaxial light) |
|||
1 | Light System Parallel contour light, 5 ring-8 sector surface light |
|||
1 | 1/2.9"CMOS digital camera | |||
3 | 0.5µm glass optical scale | |||
1 | Table for monitor | |||
1 | Documention Software manual(electronic version CHN) |
|||
1 | Optical magnification calibration board without calibration report | |||
II. | Computer System | 1 | Computer System (minimum configuration): DELL 7010MT/ I3-12100/8G*1/512G/No optical drive/DP+HDMI/Windows10H,24"display | |
1 | Software VMS3.1 2D Measuring Software (Windows O.S) |
|||
Joystick | 1 | Joystick |
Model | Measuring range(mm) | External dimensions(mm) | Machine weight(Kg) | ||||
X | Y | Z | Lx | Ly | Lz | ||
322 | 300 | 200 | 200 | 789 | 636 | 1640 | 200 |
Travel speed( mm/s) | 180 | Acceleration ( mm/s²) | 1200 |
Optical measurement accuracy(μm) | EUXY,MPE=2.5+L/200 | The unit of L is mm The measurement accuracy is based on VDI/VDE 2617-6 standard or ISO10360-7; |
Room Required Specifications | |
Room Temperature | 18~22ºC |
Temperature Gradient (in time) | 1ºC /h |
Temperature Gradient (in time) | 2ºC /24h |
Temperature Gradient (in the volume) | 1ºC /m |
Humidity Required Specifications | |
Relative humidity | 40-70% |
Electrical requirements | |
Voltage | 220V ±10% |
Frequency | 50/60 Hz |
Power consumption | 1000VA |
A good earth ground (less than 4 W) is required for reliable operation of the electrical controls. |